Direction Arrows¶
Strimma computes glucose direction (trend) locally using clinically-validated thresholds from the EASD/ISPAD 2020 consensus.
Direction Arrows¶
| Arrow | Name | Rate of Change | Meaning |
|---|---|---|---|
| ⇈ | DoubleUp | > +3.0 mg/dL/min | Rising very fast |
| ↑ | SingleUp | +2.0 to +3.0 | Rising fast |
| ↗ | FortyFiveUp | +1.1 to +2.0 | Rising |
| → | Flat | -1.1 to +1.1 | Stable |
| ↘ | FortyFiveDown | -2.0 to -1.1 | Falling |
| ↓ | SingleDown | -3.0 to -2.0 | Falling fast |
| ⇊ | DoubleDown | ≤ -3.0 | Falling very fast |
| ? | NONE | — | Insufficient data |
How Direction Is Computed¶
Strimma computes direction from your recent glucose readings rather than trusting the direction from your CGM app's notification. Here's why:
- CGM apps report direction inconsistently (different terminology, missing values)
- Local computation gives consistent, clinically-validated results
- Works across all data sources (Companion, xDrip Broadcast, Follower)
Algorithm¶
- Find a reference point — the stored reading closest to 5 minutes before the current one, at least 2.5 and at most 15 minutes old
- Average each end — average the readings in the 2.5 minutes up to the current reading, and the readings in the 2.5 minutes up to the reference point. Also average their timestamps.
- Compute delta — the change between the two averages, scaled to 5 minutes:
delta = (average_now - average_past) × 5 / minutes_between_average_timestamps - Map to direction — divide the delta by 5 to get mg/dL per minute, and compare it against the EASD/ISPAD thresholds listed above
With 1-minute readings (for example Libre 3), each end averages about 3 readings, which reduces how much a single noisy reading moves the delta and the arrow. With 5-minute readings (for example Dexcom G6), each end holds one reading, so the delta is simply the change since the previous reading. That is the same number Dexcom, xDrip+, Nightscout and AAPS show for 5-minute sensors.
The 2.5-minute window is the one Nightscout uses for its newest bucket. Nightscout doesn't divide by the real time between the averages, so on 1-minute data its delta comes out about 20% low. Strimma divides by it, so a Libre and a Dexcom show the same delta for the same rise.
Edge Cases¶
- Insufficient data: If no reading is between 2.5 and 15 minutes old, direction is "?" (NONE) and the delta is hidden
- Missed reading: If the reading 5 minutes ago is missing, the reading 10 minutes ago is used and the delta is still scaled to 5 minutes
Why Not Use the CGM App's Direction?¶
Testing against real CamAPS FX data showed a 31% mismatch rate between the CGM app's reported direction and the clinically correct direction computed from the actual readings. Common issues:
- CGM apps sometimes report "Flat" when glucose is clearly rising or falling
- Some apps use proprietary smoothing that doesn't match clinical thresholds
- Not all apps include direction in their notifications
By computing direction locally with EASD/ISPAD thresholds, Strimma provides reliable, consistent trend information regardless of the CGM app.
Clinical References¶
The direction thresholds (±1.1, ±2.0, ±3.0 mg/dL/min) come from:
- EASD/ISPAD (2020): International consensus on CGM interpretation for clinical practice
- These thresholds are widely used in diabetes technology and are the standard in most CGM systems