Direction Arrows

Strimma computes glucose direction (trend) locally using clinically-validated thresholds from the EASD/ISPAD 2020 consensus.


Direction Arrows

Arrow Name Rate of Change Meaning
⇈ DoubleUp > +3.0 mg/dL/min Rising very fast
↑ SingleUp +2.0 to +3.0 Rising fast
↗ FortyFiveUp +1.1 to +2.0 Rising
→ Flat -1.1 to +1.1 Stable
↘ FortyFiveDown -2.0 to -1.1 Falling
↓ SingleDown -3.0 to -2.0 Falling fast
⇊ DoubleDown ≤ -3.0 Falling very fast
? NONE — Insufficient data

How Direction Is Computed

Strimma computes direction from your recent glucose readings rather than trusting the direction from your CGM app's notification. Here's why:

  • CGM apps report direction inconsistently (different terminology, missing values)
  • Local computation gives consistent, clinically-validated results
  • Works across all data sources (Companion, xDrip Broadcast, Follower)

Algorithm

  1. Find a reference point — the stored reading closest to 5 minutes before the current one, at least 2.5 and at most 15 minutes old
  2. Average each end — average the readings in the 2.5 minutes up to the current reading, and the readings in the 2.5 minutes up to the reference point. Also average their timestamps.
  3. Compute delta — the change between the two averages, scaled to 5 minutes:
    delta = (average_now - average_past) × 5 / minutes_between_average_timestamps
    
  4. Map to direction — divide the delta by 5 to get mg/dL per minute, and compare it against the EASD/ISPAD thresholds listed above

With 1-minute readings (for example Libre 3), each end averages about 3 readings, which reduces how much a single noisy reading moves the delta and the arrow. With 5-minute readings (for example Dexcom G6), each end holds one reading, so the delta is simply the change since the previous reading. That is the same number Dexcom, xDrip+, Nightscout and AAPS show for 5-minute sensors.

The 2.5-minute window is the one Nightscout uses for its newest bucket. Nightscout doesn't divide by the real time between the averages, so on 1-minute data its delta comes out about 20% low. Strimma divides by it, so a Libre and a Dexcom show the same delta for the same rise.

Edge Cases

  • Insufficient data: If no reading is between 2.5 and 15 minutes old, direction is "?" (NONE) and the delta is hidden
  • Missed reading: If the reading 5 minutes ago is missing, the reading 10 minutes ago is used and the delta is still scaled to 5 minutes

Why Not Use the CGM App's Direction?

Testing against real CamAPS FX data showed a 31% mismatch rate between the CGM app's reported direction and the clinically correct direction computed from the actual readings. Common issues:

  • CGM apps sometimes report "Flat" when glucose is clearly rising or falling
  • Some apps use proprietary smoothing that doesn't match clinical thresholds
  • Not all apps include direction in their notifications

By computing direction locally with EASD/ISPAD thresholds, Strimma provides reliable, consistent trend information regardless of the CGM app.


Clinical References

The direction thresholds (±1.1, ±2.0, ±3.0 mg/dL/min) come from:

  • EASD/ISPAD (2020): International consensus on CGM interpretation for clinical practice
  • These thresholds are widely used in diabetes technology and are the standard in most CGM systems